Titanium is paramagnetic, not non-magnetic. Its volume magnetic susceptibility is roughly +150 × 10-6 (SI) at room temperature, and supplier data sheets quote a mass susceptibility around 3.2 × 10-6 cm³/g (cgs) with a relative permeability near 1.00005. Those values are small enough that a neodymium magnet will not lift a Grade 2 plate or a Grade 5 bar — but they are not zero. That distinction is not academic. It is why a magnet cannot tell you whether a bolt is Grade 2 or Grade 5, why magnetic particle inspection cannot find a crack in titanium, why an iron-contaminated surface can give a false magnetic reading, and why the word “titanium” is not an MRI safety label. This guide separates what a magnet actually proves from what it cannot, gives the measured ranges with their units and sources, and ends with a verification and MRI hand-off checklist.
Scope and verification boundary. We are a stockholder and light fabricator in Baoji, not a testing laboratory and not an MRI safety authority. This article explains published magnetic behaviour and what common tests can and cannot establish. It does not replace a certified mill test report, a qualified NDT procedure, or the device label and radiology team that decide MRI safety. Where a number appears, its unit and source basis are stated; susceptibility values differ between SI volume, cgs mass and relative-permeability conventions, so we publish the ranges rather than a single universal figure. For a critical lot, ask for a test report on that lot, not a number from an article.
| Item | Figure or finding | Why it matters | Status |
|---|---|---|---|
| Material class | Paramagnetic (weakly attracted) | Not ferromagnetic; no permanent magnetism | Industry guides / TIMET data sheet |
| Volume susceptibility (SI) | Approx +150 × 10-6 at room temperature | Positive but tiny; about 10-4 order | Industry guide, SI convention |
| Mass susceptibility (cgs) | Approx 3.2-3.3 × 10-6 cm³/g | Supplier data sheet value | TIMET-derived / supplier sheet |
| Relative permeability | Approx 1.00005 at 1.6 kA/m | Essentially 1, like a vacuum | Supplier data sheet, Ti-6Al-4V |
| Gr2 vs Gr5 by magnet | No difference a magnet can read | Both paramagnetic | Why magnet tests fail |
| Titanium vs annealed 304/316 | Both weakly non-magnetic to a shop magnet | Cold-worked 316 can turn magnetic | Stainless confusion section |
| Magnetic particle inspection | Not applicable to titanium | Use penetrant (PT) or eddy current (ET) | ASTM E1444 / E709 / ASME V |
| Iron contamination | Steel dust on surface can give a false pull | The titanium is not magnetic | Pickling / iron contamination |
| Magnetic shielding | Titanium is largely transparent to a field | A magnet can work through thin Ti | Application section |
| MRI labelling | MR Safe / MR Conditional / MR Unsafe | Titanium devices are typically MR Conditional | ASTM F2503, FDA-recognized |
How to read this table: the first four rows are the whole physics — weakly paramagnetic, never ferromagnetic. For the weight side of the 4.51 versus 4.43 figures see our weight calculator guide; here the point is that a magnet cannot read either number.
In everyday language a metal is “magnetic” if a magnet sticks to it. Titanium fails that test, so it is widely called non-magnetic. In materials physics there are three responses: ferromagnetic materials such as iron are strongly attracted and can stay magnetised; paramagnetic materials such as titanium and aluminium are attracted very weakly and do not retain magnetism; diamagnetic materials such as copper are weakly repelled.
Titanium sits in the middle group. Its susceptibility is positive but tiny, several thousand times weaker than iron, so a shop magnet sees nothing. That is why titanium works in instruments, implants and fasteners where a strong magnetic response would be a problem. It is also why people over-interpret the magnet test: “no pull” proves only “not ferromagnetic”, not “this is titanium” and certainly not “this is Grade 5”.
| Class | Examples | Behaviour in a field | Practical test result |
|---|---|---|---|
| Ferromagnetic | Iron, cobalt, nickel, martensitic stainless | Strongly attracted; can retain magnetism | Magnet sticks firmly; MPI works |
| Paramagnetic | Titanium, aluminium, platinum | Very weak attraction; no retained magnetism | Magnet will not lift it; MPI fails |
| Diamagnetic | Copper, gold, silver | Very weak repulsion | Also invisible to a shop magnet |
| Titanium in practice | Paramagnetic, chi about 10-4 SI | Thousands of times weaker than iron | “Non-magnetic” is a practical label |
Magnetic susceptibility is reported in different conventions, and mixing them is the fastest way to publish a wrong number. Volume susceptibility in SI is dimensionless; mass susceptibility in the older cgs system carries cubic centimetres per gram; relative permeability is a ratio to free space and depends on the applied field at which it is measured. We print all three with their basis rather than force them into one figure.
| Quantity | Typical published value | Unit / basis | Source basis |
|---|---|---|---|
| Volume susceptibility (SI) | Approx +150 × 10-6 | Dimensionless, SI volume | Industry guide, room temperature |
| Mass susceptibility (cgs) | Approx 3.2-3.3 × 10-6 cm³/g | cgs mass basis | Supplier / TIMET-derived data sheet |
| Relative permeability (mu r) | Approx 1.00005 at 1.6 kA/m | Dimensionless, applied field stated | Supplier data sheet, Ti-6Al-4V |
| Grade 2 vs Grade 5 | Nearly the same chi | Both paramagnetic | Why a magnet cannot sort grades |
| Iron / steel reference | Orders of magnitude higher | Ferromagnetic | Contrast only; not a Ti value |
These ranges come from industry guides and supplier data sheets, including a Ti-6Al-4V sheet quoting 3.30 × 10-6 cgs/g and a relative permeability of 1.00005 at 1.6 kA/m. They are orientation values, not our own measurements and not a certificate. If your specification carries a magnetic limit, ask a test house for a report on your actual lot and state the unit on the drawing. Our EN 10204 3.1 certificate page shows the document side of material identity.
If magnetism or grade identity matters on your order, put the verification method on the enquiry rather than relying on a magnet. These are the fields that get a meaningful answer.
|
What to send us for an exact titanium quote Copy these nine lines into your enquiry. A complete spec means a firm price on the first reply, not a second email asking for the grade. |
||||||||||||||||||||
|
||||||||||||||||||||
|
Ready? Copy the nine lines above into the Requirement Details box on our Request a Quotation page and we will come back with a firm price and a dispatch date within 24 hours. |
The most persistent shop shortcut runs like this: if a bolt is not magnetic and gives white sparks, it must be Grade 5 titanium. Both halves of that test fail on the same point. Grade 2 and Grade 5 are both paramagnetic, so the magnet cannot separate them; and every titanium grade produces a bright white spark on a wheel because the spark is the exothermic oxidation of titanium itself, not of the 6% aluminium and 4% vanadium that define Grade 5.
A counterfeit Grade 2 bolt polished to a bright finish passes both“tests” while carrying roughly a third of the yield strength of a real Grade 5 fastener. The consequence is not academic in a suspension or engine mount. This is the same lesson our traceability and PMI guide teaches from the other side: documents prove what was written, elemental analysis proves what is in the metal, and neither is replaced by a garage test.
| Test | What you observe | What it actually proves | What it misses |
|---|---|---|---|
| A neodymium magnet | No visible pull on clean Ti | Confirms “not ferromagnetic” only | Cannot identify grade or alloy content |
| White spark on a wheel | All Ti grades spark white | Confirms the element titanium | Cannot see 6% Al / 4% V of Grade 5 |
| Visual colour / finish | Gr2 and Gr5 look alike | No compositional information | Polishing can fake a higher grade |
| Density check | 4.51 vs 4.43 g/cm³ | Separates CP from Ti-6Al-4V in bulk | Needs accurate volume; see weight guide |
| XRF / spark OES | Reads elemental chemistry | The recognized PMI method | Surface-limited; see traceability guide |
| Mill test certificate | Chemistry and mechanicals per heat | Documentary proof | Must be tied to the physical heat number |
Austenitic stainless steels such as 304 and 316 are also essentially non-magnetic in the annealed condition, so a magnet does not separate them from titanium. There is one asymmetry worth knowing: thread rolling or cold heading can transform part of cold-worked 316 from austenite to martensite, which is ferromagnetic, so the threads of a stainless fastener may pull weakly while the head does not. Titanium stays paramagnetic under normal deformation.
That can flag a stainless fake, but it still does not certify titanium: a clean-looking, non-magnetic fastener could be CP titanium, Grade 5 titanium, annealed 316 or another alloy. Use chemistry. For the specification boundary between common grades see our grade equivalence guide, and for the fastener-specific failure modes see the titanium fastener page.
Here is the failure-condition chapter. A magnetic reading on a part sold as titanium is a signal to investigate, not a reason to rewrite the physics. In almost every case the pull comes from something other than the titanium matrix: transferred steel, a hidden insert, or a part that was never titanium at all.
| Situation | What causes the pull | What to do | Boundary |
| Carbon-steel tool transfer | Steel particles embedded in the surface are magnetic | Clean and re-check; the Ti is not magnetic | Iron contamination guide |
| Steel insert under a Ti shell | A hidden core or screw gives a pull | Check the whole assembly, not the label | Assemblies, not just material |
| Cold-worked 316 masquerading as Ti | Thread rolling induces martensite; threads pull weakly | Wrong material, not magnetic titanium | Magnet can flag a fake, not certify Ti |
| Elevated iron pick-up | Off-spec Fe or contamination can add a slight response | Verify by chemistry, not by feel | Supplier guide caution |
| Strong, retained magnetism | Titanium does not retain magnetism | Not titanium, or contaminated / assembled | Qualitative boundary only |
Iron contamination deserves special mention because it is invisible and easy to blame on the metal. Steel tooling, wire brushes and grinding dust can leave ferromagnetic particles embedded in a titanium surface; a magnet then grabs those particles. Our pickling and iron contamination guide explains how iron on titanium is handled on the surface side. The beam-to-beam contrast with a genuinely different failure mode is in the cavitation and erosion guide — one is surface chemistry, the other is surface mechanics.
Magnetic particle inspection magnetises a part and looks for iron particles collecting at leakage fields where a crack interrupts the field. It is fast and sensitive on steel, which is why shops reach for it. It cannot work on titanium, because there is no meaningful magnetic field in the part to interrupt.
ASTM E1444 and E709, ASME Boiler and Pressure Vessel Code Section V Article 7 and the ISO 9934 series restrict the method to ferromagnetic materials and explicitly name penetrant testing or eddy current testing for titanium, aluminium and austenitic stainless steel. If your drawing or purchase order calls out MPI on a titanium weld, that is a specification error to fix before inspection, not a technique to work around. Our weld NDT and WPS guide covers the documentation side of that call-out.
| Method | Material requirement | Use on titanium | Status |
|---|---|---|---|
| Magnetic particle inspection (MPI / MT) | Ferromagnetic materials only | Not applicable to titanium | ASTM E1444 / E709, ASME V Art. 7 |
| Liquid penetrant testing (PT) | Any non-porous surface | Surface-breaking cracks in Ti | Primary surface method for Ti |
| Eddy current testing (ET) | Electrically conductive metals | Surface / near-surface in Ti; also sorting | Works on titanium |
| Ultrasonic testing (UT) | Most metals | Internal flaws in Ti | Code-dependent technique |
| Radiography (RT) | Most materials | Volumetric flaws | Thickness / orientation limited |
The companion misconception runs the other way: somebody specifies titanium to block or shield a magnet. With a relative permeability near one, titanium is essentially transparent to a magnetic field. A magnet will attract through a thin titanium sheet, which is useful when a designer wants a non-corroding, non-reactive window over a magnet in a clasp, sensor or sealed assembly.
What titanium offers in that assembly is low weight, corrosion resistance and a weak magnetic signature, not shielding. If shielding is the requirement, that is a different material and a design problem for a competent engineer. The same “know what the material does not do” discipline applies to heat: our service temperature guide separates the oxidation limit from the strength limit just as this guide separates paramagnetic from ferromagnetic.
If grade identity or a low-magnetic requirement is on your drawing, send grade, form and verification method through our request a quotation page. A named engineer will confirm in writing which stock and certificates we can supply within 24 hours, and which checks belong to your lab or NDT provider.
This is the highest-consequence misunderstanding in the whole topic. Saying a material is titanium says nothing about whether a device made from it can enter an MRI scanner. ASTM F2503, a practice recognized in FDA guidance, defines a marking system with three categories, and it classifies the finished device under specific conditions rather than the raw material.
Under F2503, MR Safe means no known hazard in all MR environments and describes an item that is nonconducting, nonmetallic and nonmagnetic, such as a plastic petri dish. A titanium implant is metallic and conductive, so it does not qualify; most titanium implants are labelled MR Conditional, meaning no known hazard only under stated conditions of field strength, RF exposure, SAR and position. MR Unsafe covers known hazards such as ferromagnetic scissors.
| Label / test | Meaning | What it requires | Titanium relevance |
|---|---|---|---|
| MR Safe | No known hazard in all MR environments | Nonconducting, nonmetallic, nonmagnetic (e.g. plastic dish) | A titanium device is not MR Safe |
| MR Conditional | No known hazard under stated conditions | Field strength, SAR, scan time, position specified | Most titanium implants sit here |
| MR Unsafe | Known hazard in all MR environments | E.g. ferromagnetic scissors | Do not enter the MR room |
| F2052 | Magnetically induced displacement force | Tests translation in the static field | One of the F2503 referenced tests |
| F2119 | MR image artifacts from passive implants | Measures signal void / distortion | Why Ti can still blur an image |
| F2182 | RF-induced heating near passive implants | Tests temperature rise during scan | Loop geometry matters, not just grade |
The referenced test methods explain why a device cannot be cleared by material name alone. F2052 measures displacement force in the static field, F2119 measures image artefacts such as signal void around a passive implant, and F2182 measures radio-frequency heating; the last depends on geometry and conductive loops, not just the alloy. A Titanium Plate can still distort an image and a device with a loop can still heat. The label from the device manufacturer and the radiology team govern; a material supplier does not.
A material supplier can describe magnetic behaviour and state what a certificate covers. It cannot certify your inspection procedure or clear your device for a scanner.
| Assumption | Why it fails | What to do instead |
|---|---|---|
| “A magnet does not stick, so it is Grade 5 titanium” | Gr2 and Gr5 are both paramagnetic; annealed 304 is too | Verify by chemistry and MTC |
| Using the white spark test as grade proof | All Ti grades spark white; Al and V are invisible to it | Use XRF / OES / certified chemistry |
| Specifying magnetic particle inspection on Ti | MPI only works on ferromagnetic material | Call out PT or ET in the procedure |
| Calling a titanium implant “MR Safe” | F2503 requires non-metallic and non-conducting for that label | Use the device maker MR Conditional label |
| Accepting a magnetic reading on a Ti part at face value | Steel transfer or a hidden insert is the likely cause | Clean, inspect the assembly, then verify |
| Pasting a susceptibility number into a contract | Values differ by unit and source; not a lot certificate | Ask for a test report on the actual lot |
The last row governs the rest. A susceptibility figure is a material property published under a unit convention; it is not a lot acceptance value. A magnet is a qualitative screen for ferromagnetism, not a grade test. MRI clearance is a device-level, condition-specific decision. Our certifications page documents what we can state about product supply; it is not an NDT procedure or an MR label.
These fields are not a test procedure. They are the questions that let a supplier, a QA lab, an NDT Level III and, where relevant, a radiology team answer the same problem without talking past each other.
| # | Field | Example entry |
|---|---|---|
| 1 | Exact grade and spec | Gr2 ASTM B265 / Gr5 ASTM B348 / other |
| 2 | Product form | Plate / bar / wire / fastener / machined part |
| 3 | Verification needed | Grade identity / chemistry / magnetism response |
| 4 | Certificate basis | EN 10204 3.1 MTC, heat number traceable |
| 5 | NDT method called out | PT / ET / UT for Ti; not MPI |
| 6 | Surface condition | As-machined / pickled / passivated; iron-free |
| 7 | Low-magnetic requirement | Needed yes/no; acceptable susceptibility limit |
| 8 | Measuring method if specified | SI volume / cgs mass / permeability; test house |
| 9 | Assembly contents | Any steel inserts, screws or cores present |
| 10 | MR environment (if medical) | Field strength, SAR, conditions from device label |
| 11 | Responsible authority | QA lab / NDT Level III / radiology / device maker |
| 12 | Acceptance document | Which report or label is contractually binding |
Lines 3, 5 and 10 carry most of the risk: defining what must be verified, naming PT or ET instead of MPI, and taking MR conditions from a device label rather than from the word titanium. Our aerospace fastener bar page and Grade 5 bar page are the product forms where grade identity and documentation most often come together.
Choose commercially pure titanium such as Grade 2 plate when corrosion resistance, formability and the lowest practical magnetic response matter more than maximum strength. CP grades are paramagnetic like the alloys, with no retained magnetism.
Choose Grade 5 or Ti-6Al-4V such as Grade 5 bar when strength-to-weight drives the design. Do not choose it because a magnet “feels different”; it will not. Confirm the grade by chemistry and certificate, not by a bench test. For the strength-versus-fatigue context see the fatigue and surface integrity guide.
Choose a testing laboratory when a specification states a susceptibility limit, when a magnet gives an unexpected pull, or when a critical lot needs positive material identification. XRF, spark OES and a heat-numbered certificate are the recognized route; a lab report is what an auditor accepts.
Choose the device label and radiology team for anything MRI-related. Neither a supplier nor this article can turn a material name into an MR Safe or MR Conditional clearance. The same boundary applies to strength at temperature, which the heat treatment guide and service temperature guide treat as separate questions.
We are a stockholder plus light fabricator in Baoji, the centre of China’s Titanium Valley. We will not sell a magnet test as grade verification, but we can make the product and document side clear:
Start with Grade 2 plate, Grade 5 bar, aerospace fastener bar or nickel alloy stock; each page lists the grades and certification that matter for that product.
|
Get a firm titanium quote within 24 hours Same-day dispatch from ~800 t of Baoji stock · cut-to-size and waterjet blanks · free drawing review by our engineering team · EN 10204 3.1 MTC with every lot. |
Request my quote →
No obligation. A named engineer replies, not an auto-responder. |
|
EN 10204 3.1 MTC on every lot · ISO 9001 quality system · ASTM / ASME / GB standard literacy · Third-party inspection (SGS / BV) on request · Mill heat & lot traceability · Export packing & full documentation |
Strictly, yes but only very weakly. Titanium is paramagnetic: it is attracted to a strong field but far too weakly for an ordinary or neodymium magnet to lift, and it retains no magnetism after the field is removed. Industry guides put the room-temperature volume susceptibility around +150 times ten to the minus six in SI units, and supplier data sheets quote a mass susceptibility near 3.2 times ten to the minus six cubic centimetres per gram with a relative permeability around 1.00005. In shop language titanium is treated as non-magnetic, but in physics terms it is not zero.
No. Both commercially pure titanium and Ti-6Al-4V are paramagnetic and their susceptibility is nearly the same, so a hand magnet cannot tell them apart. It also cannot reliably separate titanium from annealed 304 or 316 stainless steel, which are similarly non-magnetic to a shop magnet. To confirm a grade, use elemental chemistry by XRF or spark OES together with a mill test certificate tied to the heat number; the magnet only rules out strongly ferromagnetic material.
Magnetic particle inspection needs a ferromagnetic part that can hold a magnetic field, so leakage at a crack attracts the iron particles. Titanium is paramagnetic and cannot be magnetised enough for the method to work. ASTM E1444 and E709, ASME Section V Article 7 and ISO 9934 all restrict MPI to ferromagnetic materials and name penetrant testing or eddy current testing for titanium, aluminium and austenitic stainless steel. For an internal flaw, ultrasonic or radiographic testing is used.
Usually not MR Safe in the ASTM F2503 sense. That standard defines MR Safe as no known hazard in all MR environments and describes an MR Safe item as nonconducting, nonmetallic and nonmagnetic, such as a plastic dish. A titanium device is metallic and conductive, so most titanium implants are labelled MR Conditional, meaning no known hazard only under stated conditions such as a given field strength, SAR and scan time. The device manufacturer label and the radiology team decide; the material name alone is never enough.
The titanium is almost certainly not what the magnet is holding. Common causes are carbon-steel particles transferred from tools or grinding dust into the surface, a steel insert or screw hidden under a titanium shell, or a part that is actually cold-worked stainless steel rather than titanium. If a magnet pulls strongly or the part retains magnetism, clean and inspect the assembly and verify chemistry rather than assuming the titanium itself has become magnetic.
No. With a relative permeability near one, titanium is largely transparent to a magnetic field, so a magnet can attract through a thin titanium sheet. That is useful in magnetic clasps, magnetic sensors and sealed assemblies where a non-corroding, low-reaction window is wanted over the magnet. If you need actual shielding, titanium is the wrong material and a ferromagnetic or engineered shield is required, sized by a competent designer.
State the exact grade and product form, whether you have a genuine low-magnetic requirement and any acceptance limit, and which method will be used to measure it and report the unit, because susceptibility numbers differ between SI volume, cgs mass and relative permeability. Ask for an EN 10204 3.1 certificate with a traceable heat number, call out PT or ET rather than MPI for crack inspection, and list any steel inserts in the assembly. For a medical device, take the MR conditions from the device label and leave the safety decision to the radiology team.
Titanium is paramagnetic, not ferromagnetic and not literally non-magnetic. Its susceptibility is roughly +150 × 10-6 (SI volume), about 3.2-3.3 × 10-6 cm³/g (cgs mass), with relative permeability near 1.00005 — far too weak for a shop magnet to lift, with no retained magnetism. A magnet therefore proves only “not ferromagnetic”; it cannot separate Grade 2 from Grade 5, or titanium from annealed 304 or 316, and the white spark test fails the same way. A magnetic pull on a titanium part usually means steel contamination, a hidden insert or the wrong alloy. Magnetic particle inspection cannot be used on titanium at all: ASTM E1444 and E709, ASME Section V Article 7 and ISO 9934 direct you to penetrant or eddy current instead. Under ASTM F2503 a titanium device is typically MR Conditional, never MR Safe, and the device label and radiology team decide. Verify grades by chemistry and a heat-numbered certificate, and put the method on the RFQ. Send grade, form and verification needs through our request a quotation page and a named engineer will confirm supply within 24 hours.
Titanium Weight Calculator — Quick Reference
All values use density 4.51 g/cm³ (commercially pure and most Ti alloys). For Gr5 / Ti-6Al-4V (4.43 g/cm³) multiply by 0.982. Figures are theoretical; actual weight varies with tolerance and surface condition.
1. Plate & Sheet — weight per square metre
Formula: kg/m² = thickness(mm) × 4.51
| Thickness (mm) | 0.5 | 0.8 | 1.0 | 1.5 | 2.0 | 3.0 | 4.0 | 5.0 |
|---|---|---|---|---|---|---|---|---|
| kg/m² | 2.26 | 3.61 | 4.51 | 6.77 | 9.02 | 13.53 | 18.04 | 22.55 |
| Thickness (mm) | 6.0 | 8.0 | 10.0 | 12.0 | 15.0 | 20.0 | 25.0 | 30.0 |
| kg/m² | 27.06 | 36.08 | 45.10 | 54.12 | 67.65 | 90.20 | 112.75 | 135.30 |
Per full sheet: kg = T(mm) × width(mm) × length(mm) × 4.51 ÷ 1,000,000.
2. Round Bar — weight per metre
Formula: kg/m = diameter²(mm) × 0.00354
| Diameter (mm) | 6 | 8 | 10 | 12 | 15 | 20 | 25 | 30 |
|---|---|---|---|---|---|---|---|---|
| kg/m | 0.127 | 0.226 | 0.354 | 0.509 | 0.796 | 1.415 | 2.211 | 3.183 |
| Diameter (mm) | 40 | 50 | 60 | 80 | 100 | 150 | 200 | — |
| kg/m | 5.659 | 8.843 | 12.73 | 22.64 | 35.37 | 79.58 | 141.48 | — |
3. Tube / Pipe — weight per metre
Formula: kg/m = (OD − wall)(mm) × wall(mm) × 0.01414
|